Amended in Assembly June 15, 2015

Amended in Senate May 19, 2015

Amended in Senate April 27, 2015

Amended in Senate April 6, 2015

Senate BillNo. 658


Introduced by Senator Hill

begin insert

(Principal coauthor: Assembly Member Gordon)

end insert

February 27, 2015


An act to amend Section 1714.21 of the Civil Code, and to amend Section 1797.196 of the Health and Safety Code, relating to automated external defibrillators.

LEGISLATIVE COUNSEL’S DIGEST

SB 658, as amended, Hill. Automated external defibrillators.

Existing law exempts from civil liability any person who, in good faith and not for compensation, renders emergency care or treatment by the use of an automated external defibrillator (AED) at the scene of an emergency, except in the case of personal injury or wrongful death that results from the gross negligence or willful or wanton misconduct of the person who renders emergency care or treatment. Existing law also exempts from civil liability a person or entity that acquires an AED for emergency use, a physician who is involved with the placement of the AED, and any person or entity responsible for the site where the AED is located if specified conditions are met, including maintenance and regular testing of the AED and having a written plan that describes the procedures to be followed in case of an emergency that may involve the use of the AED.begin insert Under existing law, those specified conditions also require, when an AED is placed in a public or private K-12 school, the school principal to, among other things,end insertbegin insert ensure that the school administrators and staff annually receive a brochure, approved as to content and style by the American Heart Association or the American Red Cross, that describes the proper use of an AED, to ensure that similar information is posted next to every AED, and to designate the trained employees who are available to respond to an emergency that may involve the use of an AED during normal operating hours.end insert

This bill would provide an exemption from civil liability for a physician and surgeon or other health care professional that is involved in the selection, placement, or installation of an AED. The bill would require a person or entity, other than a health facility as defined, that acquires an AED to, among other things, comply with specified regulations for the placement of the device and ensure that the AED is maintained and tested as specified. The bill would require a building owner to annually notify the tenants as to the location of the AED units and provide information to tenants about who they can contact if they want to voluntarily take AED or CPR training, to offer a demonstration to at least one person associated with the building as to the use of an AED in an emergency, and post instructions for the use of the AED. The bill would also specify that a medical director or physician and surgeon is not required to be involved in the acquisition or placement of an AED. The bill would make related changes.

begin insert

This bill would revise the public or private K-12 school provisions described above by instead requiring, when an AED is placed in a public or private K-12 school, the school principal to ensure that the school administrators and staff annually receive information that describes sudden cardiac arrest, the school’s emergency response plan, and the proper use of an AED, by instead requiring the school principal to ensure that instructions, in no less than 14-point type, on how to use the AED are posted next to every AED, and by deleting the requirement that the school principal designate the trained employees who are available to respond to an emergency that may involve the use of an AED during normal operating hours.

end insert

Vote: majority. Appropriation: no. Fiscal committee: no. State-mandated local program: no.

The people of the State of California do enact as follows:

P2    1

SECTION 1.  

Section 1714.21 of the Civil Code is amended
2to read:

P3    1

1714.21.  

(a) For purposes of this section, the following
2definitions shall apply:

3(1) “AED” or “defibrillator” means an automated external
4defibrillator.

5(2) “CPR” means cardiopulmonary resuscitation.

6(b) Any person who, in good faith and not for compensation,
7renders emergency care or treatment by the use of an AED at the
8scene of an emergency is not liable for any civil damages resulting
9from any acts or omissions in rendering the emergency care.

10(c) A person or entity who provides CPR and AED training to
11a person who renders emergency care pursuant to subdivision (b)
12is not liable for any civil damages resulting from any acts or
13omissions of the person rendering the emergency care.

14(d) (1) A person or entity that acquires an AED for emergency
15use pursuant to this section is not liable for any civil damages
16resulting from any acts or omissions in the rendering of the
17emergency care by use of an AED if that person or entity has
18complied with subdivision (b) of Section 1797.196 of the Health
19and Safety Code.

20(2) A physician and surgeon or other health care professional
21that is involved in the selection, placement, or installation of an
22AED pursuant to Section 1797.196 of the Health and Safety Code
23is not liable for civil damages resulting from acts or omissions in
24the rendering of emergency care by use of that AED.

25(e) The protections specified in this section do not apply in the
26case of personal injury or wrongful death that results from the
27gross negligence or willful or wanton misconduct of the person
28who renders emergency care or treatment by the use of an AED.

29(f) This section does not relieve a manufacturer, designer,
30developer, distributor, installer, or supplier of an AED or
31defibrillator of any liability under any applicable statute or rule of
32law.

33

SEC. 2.  

Section 1797.196 of the Health and Safety Code is
34amended to read:

35

1797.196.  

(a) For purposes of this section, “AED” or
36“defibrillator” means an automated external defibrillator.

37(b) (1) In order to ensure public safety, a person or entity that
38acquires an AED shall do all of the following:

39(A) Comply with all regulations governing the placement of an
40AED.

P4    1(B) Notify an agent of the local EMS agency of the existence,
2location, and type of AED acquired.

3(C) Ensure that the AED is maintained and tested according to
4the operation and maintenance guidelines set forth by the
5manufacturer.

6(D) Ensure that the AED is tested at least biannually and after
7each use.

8(E) Ensure thatbegin delete a visualend deletebegin insert anend insert inspection is made of all AEDs on
9the premises at least every 90 days for potential issues related to
10operability of the device, including a blinking light or other obvious
11defect that may suggest tampering or that another problem has
12arisen with the functionality of the AED.

13(F) Ensure that records of the maintenance and testing required
14pursuant to this paragraph are maintained.

15(2) When an AED is placed in a building, the building owner
16shall do all of the following:

17(A) At least once a year, notify the tenants as to the location of
18the AED units and provide information to tenants about who they
19can contact if they want to voluntarily take AED or CPR training.

20(B) At least once a year, offer a demonstration to at least one
21person associated with the building so that the person can be
22walked through how to use an AED properly in an emergency.
23The building owner may arrange for the demonstration or partner
24with a nonprofit organization to do so.

25(C) Next to the AED, post instructions, in no less than 14-point
26type,begin delete from the manufacturerend delete on how to use the AED.

27(3) A medical director or other physician and surgeon is not
28required to be involved in the acquisition or placement of an AED.

29(c) (1) When an AED is placed in a public or private K-12
30school, the principal shall ensure that the school administrators
31and staff annually receivebegin delete a brochure, approved as to content and
32style by the American Heart Association or the American Red
33Cross, that describesend delete
begin insert information that describes sudden cardiac
34arrest, the school’s emergency response plan, andend insert
the proper use
35of an AED. The principal shall also ensure thatbegin delete similar information
36isend delete
begin insert instructions, in no less than 14-point type, on how to use the
37AED areend insert
posted next to every AED. The principal shall, at least
38annually, notify school employees as to the location of all AED
39units on the campus.begin delete The principal shall designate the trained
40employees who shall be available to respond to an emergency that
P5    1may involve the use of an AED during normal operating hours.
2As used in this subdivision, “normal operating hours” means during
3the hours of classroom instruction and any school-sponsored
4activity occurring on school grounds.end delete

5(2) This section does not prohibit a school employee or other
6person from rendering aid with an AED.

7(d) A manufacturer or retailer supplying an AED shall provide
8to the acquirer of the AED all information governing the use,
9installation, operation, training, and maintenance of the AED.

10(e) A violation of this section is not subject to penalties pursuant
11to Section 1798.206.

12(f) Nothing in this section or Section 1714.21 of the Civil Code
13may be construed to require a building owner or a building manager
14to acquire and have installed an AED in any building.

15(g) For purposes of this section, “local EMS agency” means an
16agency established pursuant to Section 1797.200.

17(h) This section does not apply to facilities licensed pursuant
18to subdivision (a), (b), (c), or (f) of Section 1250.



O

    95